Flowtron Universal Service Manual
Testing
Issue 2
Chap 4
Sep 2009
SER0009
Page 13
Figure 30 - Inserting the Test Probe through the Rear Case
Figure 31 - Earth/Ground Test Area on Control PCB Assembly
Test Probe
Make sure the mains/power
cordset connector is pushed
securely into the pump
Mains/Power Cordset
Screw
Case IEC Cover
Note:
Insert the end of the test probe through the test hole in the rear case.
Rear Case of Pump
Approximate dimensions of test probe: Minimum length = 100 mm
Maximum diameter = 2 mm
Earth/Ground
Access Point
Control PCB Assembly
Compressor Assembly
Earth/Ground Test Pad
IEC Connector
Front Case of Pump
(inside dotted line)
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...