Page 19
Figure 87 - Communicator Software - Flash Loader Fail
10.8
On the dialog box, under Download Option, click Load with Flash Loader to start the software
download of the V3.01 Flowtron Universal software to the pump.
10.8.1
If the
Flash Loader Fail
dialog box (refer to Figure 87) is displayed when you click
Load with Flash Loader
, click
OK
:
10.8.1.1
The software download should start. If so, continue the procedure at
Para 10.9, below.
Note:
It is normal for the
Flash Loader Fail
dialog box to be displayed the
first time you click
Load with Flash Loader
.
10.8.1.2
If the
Flash Loader Fail
dialog box continues to be displayed every time
you click
Load with Flash Loader
, then carry out the procedure detailed in
Chapter 6 “Updating the software”, Page 13, “Communication Error when
Executing Commands”.
10.9
As the software is downloaded, the progress is shown in the dialog box, as follows (refer to
Figure 86):
•
The
Status
part of the dialog box shows
Loading Bank 0
(followed by
Loading Bank 1
and
then
Loading Bank 2
)
and a moving
Progress
“status bar”.
•
The
Tx Rx
box counts up from
B0 S0 OK
to
B0 S128 OK
(followed by
B1 S0 OK
to
B1 S128 OK
and then
B2 S0 OK
to
B2 S128 OK
).
10.10 The software download is complete when:
•
The
Status
part of the dialog box is empty (when the
Loading Bank 2
and the moving
Progress
“status bar” have been removed).
•
The count in the
Tx Rx
box gets to
B2 S128 OK
.
10.11 At the end of the download, the pump will automatically start a self-test sequence, as detailed in
Page 14, Table 31:
10.11.1 If the self-test sequence is successful, check the version of the software on the pump to
make sure the software has been downloaded correctly (refer to
Page 19, Section 10.11.2).
10.11.2 If the pump fails the self-test sequence, the pump is faulty and must be rejected.
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...