Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
Part number: 801 073 81
Description: Keyacid Fluorescein 019187
Supplier: Keystone Europe Ltd.
Contact:
Be sure to handle chemicals with care and to read and comply with all health and
safety advice.
15.9.5.
Serial Dilution
A three step dilution process should be used as outlined below.
Step 1
; Weigh out 0.5g Fluorescein dye powder and add to 1L deionized water in a
volumetric flask. Invert 10 times or until all powder is dissolved. This gives a stock solution
of 500mg/L.
Step 2
; Transfer 10ml of the 500mg/L stock solution into a 1L volumetric flask and top the
flask up to 1L with deionized water. Invert to mix.
This step results in a 1 in 100 dilution of the 500mg/L stock resulting in a 5mg/L stock.
Step 3
; Transfer 20ml of the 5mg/L stock from step 2 into a 1L volumetric flask. Top up to
1L with deionized water. Invert to mix.
This step results in a 1 in 50 dilution and gives you the 100μg/L FSCEIN calibration
standard required for Pt-2.
The dilute solution can be stored in a dark bottle in a refrigerator for up to five days. After
that time it must be discarded.
15.9.6.
Zero Point Calibration
To calibrate the zero point, follow these steps:
1. Pour 400mL of clean water (bottled still mineral water is recommended) into a clean
calibration cup, remove the storage cap from the pH electrode if fitted, wash the
Probe in clean water, then gently lower the Probe in all the way.
The Sleeve End
Cap must be fitted
. Activate the probe cleaning feature in order to remove any air
bubbles that may be clinging to the electrodes.
2. Switch the Aquameter
®
on and wait until the temperature and Fcein readings are
stable. If the Fcein reading is very high, there are probably air bubbles adhering to
the lenses.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal