Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
11. After Use
The AP-7000 should always be cleaned after every use.
It is advisable to clean the Probe after use with the cable attached. This will prevent
any water entering the Probe's socket and will allow any deposits to be removed
from the connector collar and shell.
The Sleeve on the AP-7000 can be removed by unscrewing to allow cleaning of the
individual electrodes. After every use, remove the protective Sleeve End Cap then unscrew
the sleeve.
With the Sleeve removed, the individual electrodes are very vulnerable, so please
handle the Probe with extreme care. If you drop it, it’s going to break!
Rinse the exposed electrodes, the inside of the Sleeve and the Sleeve End Cap with fresh,
clean water. Shake the water from inside the Sleeve, then reattach. Dry the outside of the
Probe using a soft cloth.
Remember to replace the pH/ORP storage cap after use
. Failure to do so will damage
the electrode. For more details, see
in section 13.
Never clean the Probe with solvents, alcohol or concentrated acid/alkaline based
cleaning products such as Decon 90. These products can strip the anodised finish
from the Probe and damage the plastic and rubber components. Damage caused by
the use of aggressive cleaning agents or solvents is not covered by your warranty.
TIP:
Occasional application of a smear of silicone grease (provided) to the connector O-
rings and thread, Sleeve thread, the protective Sleeve End Cap thread will make fitting and
removal of these parts easier.