Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
Once all three solutions are at a stable temperature, calibration can begin.
16.8.2.
Three-point Calibration
During three-point calibration, the AP-7000 and Aquameter
®
must remain switched on. If
the Aquameter
®
is switched off between points, the calibration process will be aborted and
must be re-started from point 1. The AP-7000's sleeve and wiper should also be removed
in order to reduce the Probe's thermal mass. To calibrate the ISE electrode follow these
steps:
Point 1.
1. Remove the Probe Sleeve and the Wiper (see section
). Remove the storage
cap from the pH electrode, wash the Probe in distilled water, dry the probe
thoroughly then gently lower the Probe in to the warm
0.5ppm
solution.
2. Switch the Aquameter
®
on and stir the probe until the temperature and F readings
are completely stable. A minimum of five minutes is recommended.
3. Ensure the temperature of the solution is between 20ºC and 40ºC (68ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5. Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6. Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:TURB
l
4:EMPTY
2:F
l
5:EMPTY
3:EMPTY
l
6:EMPTY
The Fluoride (F) electrode should have been assigned to an AUX socket when it was fitted.
Move the pointer to F then press the OK or right arrow key to select.
7.The screen will change to:
CALIBRATE F
Pt-1? [01/Jan/17]
Pt-2? [01/Jan/17]
Pt-3? [01/Jan/17]
Calibration point 1 (Pt-1) is the warm 0.5ppm point. Calibration point 2 (Pt-2) is the warm
5ppm point. Calibration point 3 (Pt-3) is the cool 0.5ppm point.