Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
15.5.6.
Zero Point Calibration
To calibrate the zero point, follow these steps:
1. Pour 400mL of clean water (bottled still mineral water is recommended) into a clean
calibration cup, remove the storage cap from the pH electrode if fitted, wash the
Probe in clean water, then gently lower the Probe in all the way.
The Sleeve End
Cap must be fitted
. Activate the probe cleaning feature in order to remove any air
bubbles that may be clinging to the electrodes.
2. Wait until the temperature and BGA-PC readings are stable. If the BGA-PC reading
is very high, there are probably air bubbles adhering to the lenses.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6.
Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:BGA-PC
l
4:EMPTY
2:EMPTY
l
5:EMPTY
3:EMPTY
l
6:EMPTY
The BGA-PC electrode should have been assigned to an AUX socket when it was fitted.
Choose that socket. Press the OK or right arrow key to select BGA-PC. The screen will
change to:
CALIBRATE BGA-PC
ZERO? [01/Jan/17]
Pt-2? [01/Jan/17]
GS Factor:01.00
Calibration point 2 (Pt-2) is the upper calibration point.
The dates shown to the right of each point are the dates of the last successful calibration.
7.
Select ZERO. The screen will change to:
© 2017 Aquaread
®
Ltd.