10.3. Using RapidCal
RapidCal calibrates EC at 2570µS/cm and the pH7.00 point simultaneously. Ideally, this
procedure should be carried out at the beginning of each day the Probe is to be used. In
addition, you should check the DO 100% calibration point and zero any optical electrodes
you may have fitted. To use RapidCal:
1. Pour 400mL of RapidCal solution into one
of the calibration cups provided as shown in
the adjacent photograph.
2. Remove the storage cap from the pH
electrode if fitted, wash the Probe in
distilled water, then gently lower the Probe
into the calibration cup.
3. If you do not have any ISE caps fitted,
activate the probe cleaning feature in order
to remove any air bubbles that may be
clinging to the electrodes. To do this, press
the MENU key and select CLEAN PROBE.
If you have ISE caps fitted, agitate the
Probe to eliminate air. Do not activate
the cleaning mechanism as the brush
will jam against the ISE caps.
4. Wait until all readings are
completely
stable. The longer you can leave the probe
to achieve thermal equilibrium before
proceeding, the better. A minimum of two minutes is recommended.
5. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF – 104ºF). The
closer to 25ºC the better.
6. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
7. Select
RapidCal.
The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until all readings are stable, then it will send the RapidCal command to
the Probe, where the calibration takes place.