12. The AP-7000 Automatic Cleaning System
The AP-7000 features a motorised cleaning mechanism. This cleaning mechanism is
designed for use during long term deployment when the AP-7000 is connected to an
AquaLogger or a third party telemetry device capable of supplying the required voltage and
current.
The cleaning cycle can be activated from the Aquameter
®
, which is useful during calibration
for removing air bubbles from the electrodes.
To activate the cleaning mechanism from the Aquameter
®
, press the
MENU
key. The Main
Menu screen will be displayed. Select the first item on the menu, 'Clean Probe'.
Clean Probe
Auto Data Logging
Calibration
Setup & Install
When activated, the centrally mounted wiper will rotate to each occupied electrode position
and carry out a brief scrubbing action before returning the its park position.
If there is not enough power available from the meter or if the meter-probe cable is
excessively long (which causes a voltage drop), a 'Wiper Jammed' message will be
displayed.
This message will also be displayed if the wiper hits an obstacle which prevents it
completing it's cleaning cycle. If an obstacle is hit and the wiper jams, it will reverse
direction and return to its park position.
12.1. Wiper Removal and Replacement
Occasional cleaning or replacement of the wiper brushes may be required due to fouling.
This is a simple job and requires no tools.
First remove the sleeve then pull the stainless-steel spring pin out of the central shaft.