27.7. Cleaning
After use, rinse the Flowcell thoroughly with fresh water. To remove stubborn deposits,
scrub the inside of the Flowcell with a bottlebrush and non-abrasive detergent, then rinse
thoroughly.
Never clean the Flowcell with concentrated acid or alkaline based cleaning products
such as Decon 90. These products can strip the anodised finish from the Flowcell
and damage the plastic components.
27.8. Flowcell Troubleshooting
Problem
Cause / Remedy
DO readings are abnormally
high or are fluctuating wildly.
✔
Aeration of sample water. Check all joints for air
leaks. Reduce flow rate to avoid cavitation.
Turbidity or other optical
electrode readings are
abnormally high or are
fluctuating wildly.
✔
Air bubbles adhering to the Turbidity Electrode
lenses. Agitate Flowcell to dislodge.
✔
Aeration of sample water. Check all joints for air
leaks. Reduce flow rate to avoid cavitation.
✔
Optical electrode not zeroed in the Flowcell. Fill
the Flowcell with clean water and Zero the optical
electrode.