6. Taking Measurements
The AP-7000 includes a pH/ORP electrode, which is kept moist by a storage cap. Remove
the storage cap by pulling the red lanyard marked ‘Remove Before Use / Replace After
Use’ straight down.
Do not use a twisting motion to remove or replace the cap as this
can unscrew the electrode from the Probe body.
Rinse any salty deposits from the
pH/ORP electrode with fresh water.
Fit the protective Sleeve End Cap into the end of the Probe sleeve. Switch the Aquameter
®
on and immerse the AP-7000 in the sample water, making sure that the water level covers
the minimum immersion depth groove halfway up the Probe sleeve.
TIP:
Occasional application of a smear of silicone grease or similar lubricant to the
protective Sleeve End Cap thread will make fitting and removal of the Cap easier.
If the AP-7000 is connected correctly, the meter will read the Probe’s serial number and
model number, then will automatically configure itself to display only those readings the
current AP-7000 is capable of taking. Initial Probe readings will be displayed on the meter’s
screen along with the current GPS status. The initial data screen for the GPS Aquameter
®
in conjunction with the AP-7000 is shown below.
TEMP:018.5
°
C
ORP:0415.2 mV
pH:06.48
GPS:Acquiring
Left/right arrows at the bottom corners of the screen indicate further data screens are
available. To access these screens, simply press either the left or right arrow keys. Any
value that is out of range or unavailable will be displayed as dashes. The other four screens
available with the standard AM-200/AP-7000 combination are shown below.
DO:098.7%
EC:6541µS/cm
TDS:3271mg/L
GPS:3D Pos
SAL:03.57 PSU
SSG:01.3σt
GPS:3D Pos
BARO:1013mb
DEPTH:01.75m
Hit [OK] to zero
GPS:3D Pos
© 2017 Aquaread
®
Ltd.
www.aquaread.com