Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
The dates shown to the right of each point are the dates of the last successful calibration.
8. Select Pt-1. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Output:348mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage output from the ISE electrode in
millivolts (mV). This value is stored in the Probe's memory and can be recalled at any time.
See section
Calibration Data Storage and Retrieval.
Press OK then ESC repeatedly to
return to normal reading mode.
Point 2
1. Remove the probe from the 0.5ppm solution and wash thoroughly in deionised
water. Dry the probe then gently lower it into to the warm
5ppm
solution.
2. Stir the probe until the temperature and F readings are completely stable. A
minimum of five minutes is recommended.
3.
Ensure the temperature of the solution is within 1ºC of the previous 0.5ppm
calibration point.
If the solution is warmer or cooler than this, calibration will fail.
4. Referring to steps 4-7 above, select Pt-2 and press OK.
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up as shown above. If the calibration is
successful, the counter will reach 100% and the calibration report screen will be displayed.
If the temperature of the 5ppm solution is more than 1ºC different from the Pt-1 calibration
temperature, an OUT OF TEMP RANGE calibration error will be reported. If this happens,
adjust the temperature and try again.
Point 3
1. Remove the probe from the 5ppm solution and wash thoroughly in deionised water.
Dry the probe then gently lower it into to the
cool 0.5ppm
solution.
2. Stir the probe until the temperature and F readings are completely stable. A
minimum of fifteen minutes is recommended.
3.
Ensure the temperature of the solution is at least 10ºC cooler than the
previous 5ppm calibration point.
If the solution is too warm, calibration will fail.
4. Referring to steps 4-7 above, select Pt-3 and press OK.