Calibration
RapidCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6. Select
DO/EC
. The screen will change to:
Calibration
DOZero?[01/Jan/17]
DO100%?[01/Jan/17]
EC2570?[01/Jan/17]
The dates shown to the right of the screen are the dates of the last successful calibration.
7. Select DOZero. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Output:4.4
Calibrating
100%
Press [OK]
The top line will display a value which represents the health of the luminophore. This value
should be between 3.5 and 4.5 (at 25°C). If the value returned is less than 3.5, the Optical
DO Cap should be replaced.
This value is stored in the Probe's memory and can be recalled at any time. See section
Calibration Data Storage and Retrieval.
Press OK then ESC repeatedly to return to
normal reading mode.