When the Socket Assignment option has been selected, the following screen will be
displayed.
The numbers 1 – 6 represent the AUX socket numbers.
SOCKET ASSIGNMENTS
1:EMPTY
l
4:EMPTY
2:EMPTY
l
5:EMPTY
3:EMPTY
l
6:EMPTY
Using the up and down arrow keys, select the AUX socket you wish to assign then move
the cursor to the right by pressing the right arrow key. When the cursor has moved to the
right of the AUX socket number, use the up and down arrow keys to select the appropriate
electrode type.
The tables below show the available electrode options and the selection that should be
made on this screen:
AP-7000 Optical Electrodes
Electrode Part No.
Function
Aquameter
®
Selection
7000-TURB
Turbidity
TURB
7000-CPHYLL
Chlorophyll
Cphl
7000-BGA-PC
Phycocyanin (Blue-Green Algae PC)
BGA-PC
7000-BGA-PE
Phycoerythrin (Blue-Green Algae PE)
BGA-PE
7000-RHOD
Rhodamine WT Dye
Rhod
7000-FSCEIN
Fluorescein Dye
Fcein
7000-REFOIL
Refined Oil
OIL
7000-CDOM
CDOM/FDOM
CDOM
AP-5000/7000 ISE Electrodes
Electrode Part No.
Function
Aquameter
®
Selection
5000/7000-AMM
Ammonium/Ammonia
NH4
5000/7000-CHL
Chloride
Cl
5000/7000-FLU
Fluoride
F
5000/7000-NIT
Nitrate
NO3
5000/7000-CAL
Calcium
Ca2
When the desired electrode type is showing, move the cursor back to the left of the socket
number then press OK to send the selection to the AP-7000. The socket assignments are
stored in the AP-7000.
If you press the ESC key whilst in this screen, any changes you have made will not be
transferred to the AP-7000.
Please note: changing an AUX Socket assignment will clear all the calibration data
for that socket.
If you subsequently remove an electrode, be sure to set the socket assignment back to
EMPTY.