22. Appendix 2. Fitting AUX Electrodes
There are two types of AUX Electrodes designed for use with the AP-7000: Optical
Electrodes and ISE Electrodes. These and can be fitted to any of the AUX sockets. The six
AUX sockets on the AP-7000 should be populated in numeric order starting with AUX1. If
you are fitting a mix of Optical and ISE Electrodes, fit the Optical Electrodes first.
Installing AUX Electrodes
First, remove the blanking plug from the AUX socket that you want to use on the AP-7000.
To remove the blanking plug and subsequently tighten the AUX Electrode, use the red
lanyard that is attached to the pH/ORP storage cap as a belt wrench as shown below.
Apply a small amount of silicone grease (supplied with the AP-7000) to the threaded
section and the O-ring of the AUX Electrode (see photograph).
ENSURE NO GREASE IS APPLIED TO THE GOLD CONTACTS
. Using a clean cloth or
tissue paper, polish the gold contacts ensuring they are completely clean. Carefully insert
the electrode into the AUX socket and tighten firmly until the O-ring is completely
compressed.
Apply a small amount of silicone grease (supplied with the AP-7000) to the threaded
section and the O-ring of the AUX Electrode (see photograph).
ENSURE NO GREASE IS APPLIED TO THE GOLD CONTACTS
. Using a clean cloth or
tissue paper, polish the gold contacts ensuring they are completely clean. Carefully insert
the electrode into the AUX socket and tighten firmly until the O-ring is completely
compressed.