![Aquaread Aquaprobe AP-7000 Instruction Manual Download Page 41](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-7000/aquaprobe-ap-7000_instruction-manual_2960645041.webp)
Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
Calibration
RapidCal
DO 100%
Full Cal
6. Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
7. Select
pH/REDOX(ORP)
. The screen will change to:
pH 7.00?[01/Jan/17]
PH 4.01?[01/Jan/17]
PH 10.0?[01/Jan/17]
ORP+250?[01/Jan/17]
The dates shown to the right of the screen are the dates of the last successful calibration.
8. Select pH7.00. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Offset: -1.2mV
Calibrating
100%
Press [OK]
The top line displays the voltage offset from zero for the pH electrode in +/-millivolts (mV). If
this offset goes /-25mV at 25
°
C, the pH electrode should be serviced.
This value is stored in the Probe's memory and can be recalled at any time. See section
Calibration Data Storage and Retrieval.
Press OK then ESC repeatedly to return to
normal reading mode.
© 2017 Aquaread
®
Ltd.
www.aquaread.com
Page 41 of 143