Portable Test Equipment
User’s Guide
www.appliedp.com
PTE-UGFW56a-EN
99
SYST:SOFT:ENAB?
Request for Soft Generate general enable/disable parameter. Answer is 0 (=disabled) or 1
(=enabled).
SYST:SOFT:ENAB
{ enable(0|1) }
Setting of Soft Generate general enable/disable parameter. Parameter
enable
can be 0 (=disable) or
1 (=enable).
SYST:SOFT:DEL?
Request for Soft Generate Delay parameter. Answer is time in milliseconds.
SYST:SOFT:DEL
{ delay(0-65535) }
Setting of Soft Generate Delay parameter. Parameter
delay
is time in milliseconds.
Soft Generate Delay parameter is duration of one step of ON/OFF transitions (generation start or
stop).
SYST:BC:ENAB?
Request for Back Control general enable/disable parameter. Answer is 0 (=disabled) or 1 (=enabled).
SYST:BC:ENAB
{ enable(0|1) }
Setting of Back Control general enable/disable parameter. Parameter
enable
can be 0 (=disable) or 1
(=enable).
SYSTem:BC:REPeat?
Request for Back Control repeat parameter. Answer
repeat
can be: 0 (=infinite), 1 (=once), 2 (=twice).
SYSTem:BC:REPeat { repeat(0|1|2) }
Setting of Back Control repeat parameter. Parameter
repeat
determines how many times Back
Control runs: 0 (=infinite), 1 (=once), 2 (=twice)
SYST:SYNC:ENAB?
Request for state of Input Synchronization. Answer is 0 (=disabled) or 1 (=enabled).
SYST:SYNC:ENAB
{ enable(0|1) }
Setting of state of Input Synchronization. Parameter is 0 (=disable) or 1 (=enable).
GEN:CONT:ERR?
Request for actual state of current system errors and voltage systems errors. Return value is a 16-bit
word holding information about errors in system.
Bit #
15
14 13
12 11
10
9
8
7
6
5
4
3
2
1
0
Value
0
0
U
3
I
3
U
2
I
2
U
1
I
1
0
0
0
0
0
0
0
0
Bit meanings
Ix
error in current system in phase Lx (indicated also on panel by deactivated Signal
LED)
Ux error in voltage system in phase Lx (indicated also on panel by deactivated Signal
LED)
This error is indicated while load point generation is active. If this generation is done without problems
in current or voltage system then this indication disappears automatically.