Phase Scale
................ Phase scale with values 0-360 or 0-
360/n …
n= harmonic number; in
second case will be phase values in these ranges: 2
nd
harmonic 0-
180
°, 3
rd
harmonic 0-120
°, 4
th
harmonic 0-90
° etc.
ADDITIONAL INFO
................. Settings of special measurement
Power of harm.
............ Calculation and display of Active Power of higher harmonics (for 2
nd
– 31
st
harmonic) on the Load Values screen:
ON
(enabled) or
OFF
(disabled). Sign of the number determines direction of energy flow:
positive means energy consumption, negative energy delivering. If
the sign of this quantity (Ph) is different than the sign of Active power
(P), it means that higher harmonics are in opposite direction against
the total power.
Curr. phase seq.
......... Display of Current Phase Sequence on the Load Values screen:
ON
(enabled) or
OFF
(disabled)
Lead/Lag indic.
........... Enable of Indication of leading or leading in the row cos
φ on Load
Values screen. Indication is presented as small red triangle at the
corners of each channel L1, L2, L3, while up is LAG, down is LEAD.
Meas. Settings screen 5/6
* P91
Meas. Settings screen 6/6
* P201
VOLTAGE AMPLITUDE
.......... Settings of voltage drop warnings
Check
.......................... Enable
(ON)
or disable (
OFF)
checking of voltage amplitude value
Threshold
.................... Value in %, relative to nominal voltage of selected meter to be tested
CURRENT AMPLITUDE
........... Settings of current drop warnings
Check
.......................... Enable
(ON)
or disable (
OFF)
checking of current amplitude value
Threshold
.................... Value in %, relative to nominal current of selected meter to be tested
SHOW MEAS. VALUES IN DB
Settings of measured values in database
Prim. only
.................... When
OFF
the database will contain only result values of the
secondary side of measurement. Respectively the side of
measurement where it was carried out. When
ON
the database will
contain only result values of the primary side of measurement, if the
measurement was carried out on the secondary side, and there are
values of CT and VT not equal to 1.
Result values here are values from Meter Error Test and Energy Test
– voltages and currents.
PHASE PRESENTATION
......... Settings of phase presentation for the whole device
Range
.......................... setting the normalization of phase values; whether it is
(0
– 360 °)
or (
±180°)