Portable Test Equipment
User’s Guide
PTE-UGFW56a-EN
www.appliedp.com
82
SYST:ENER:IMP:K {<factor>}
Setting of the multiplication constant of the device’s impulse output which generates the impulse
signal proportional to the measured energy. Parameter <factor> is the multiplication constant of the
impulse output in the unit of 1/kWh or 1/kVAh.
SYST:ENER:IMP:STAT?
Request for the state of the device impulse output. Reply is number:
0 … state is OFF
1 … state is ACTIVE POWER
2 … state is REACTIVE POWER
3 … state is FIXED
4 … state is APPARENT POWER
SYST:ENER:IMP:STAT {0|1|2|3|4} | {OFF|ACTIVE|REACTIVE|FIXED|APPARENT}
Setting of the device impulse output state. Parameter is state code (0 to 4 listed in previous
command) or state name.
SYST:ENER:IMP:FIX?
Request for the constant frequency value (which is generated if the impulse output is in the state
FIXED). Reply is the text string containing a real number with the value of this frequency (e.g.
“+1.03”). This value is in a Hz unit.
SYST:ENER:IMP:FIX {<freq>}
Setting of the constant frequency value (which is generated if the impulse output is in the state
FIXED). Parameter <freq> is the frequency value in Hz unit.
SYST:ENER:IMP:MOD?
Request for the state of the device impulse output. Reply is number:
0 … state is FIXED
1 … state is ACTIVE POWER
2 … state is REACTIVE POWER
3 … state is APPARENT POWER
SYST:ENER:IMP:MOD {<mode>}
Setting the state of device’s impulse output. Parameter is number:
0 … state is FIXED
1
… state is ACTIVE POWER
2 … state is REACTIVE POWER
3 … state is APPARENT POWER
SYST:ENER:IMP:AUX?
Request for the state of AUX output. The same parameter as described in Measurement Settings,
page 53. Answers are 0 = OFF, 1 = ON.
SYST:ENER:IMP:AUX {<auxiliary>}
Setting of the state of AUX output. The same parameter as described in Measurement Settings, page
53. Parameter can be 0 = OFF, 1 = ON.
SYST:ENER:DOS:INIT
Initialization of dosage function.
SYST:ENER:DOS:STAR
Start (or initialization) of dosage function.