Portable Test Equipment
User’s Guide
www.appliedp.com
PTE-UGFW56a-EN
25
Constant
................... Meter constant
– from database
(
Meter DB
) or user defined (
User def.
)
User Constant
..........
User defined Constant …
Usable only if Constant = “User def.”
Constant Unit
........... User defined Constant Unit
– i/kWh, Wh/i, Ws/i, …
Usable only if Constant =
“User def.”
Constant Type
.......... User defined Constant Type
–
secondary, primary
;
Usable only if Constant =
“User def.”
Add Test screen 3/4
* P129
Add Test screen 4/4
* P130
Sample Type
............. Sample Type -
Int. Time, Impulses
Integration Time
....... Integration time in seconds;
Usable only if Sample Type = “Int. Time”
No. of impulses
........ Pulses count;
Usable only if Sample Type = “Impulses”
No. of Samples
......... How many samples to calculate resulting error
Energy Dose
............. How much energy to dose for Dial (Register) Test; Unit is kWh;
Usable only if
Type = “Dial Test”
Reg. Channels
.......... Which channels to register
– L1, L2, L3
Method
...................... Mode of energy registration
– from database (
Meter DB
) or user defined
(
User def.
)
User Method
............. User definable mode of energy registration
– Algo0…Algo4
Usable only if
Method = “User def.”
User Sense
............... Energy register type -
Import, Export
(see chapter 5.7.1 Measurement
Frequency
................. Frequency during test -
Meter DB, User def.
User. Freq.
................ User definable frequency;
Usable only if Frequency
= “User def.”
Back Control
............ Enabling of BC feature
–
ON, OFF, No change
Readout reg.
............. Register address within electricity meter, where dialer value is stored and
which will be verified during dial test. If not filled, then during dial test it is
necessary to fill fields
Counters #1
(at the beginning of dial test) and
Counters #2
(at the end of dial test) on (blue) screen Dial test, see chapter
Min. (Max.) Pulses
.... not implemented yet (for future use)
Show message
......... not implemented yet (for future use)
Tolerance
.................. Accuracy class related to given test -
Meter DB, User def.
User Tolerance
......... User definable accuracy class;
Usable only if Tolerance = “User def.”
Toler. Multiplier
........ Multiplication factor of Tolerance (for easy increasing / decreasing tolerance
limits)