SYST:ENER:DOS:ENER:K?
Request for energy (in kWh or kVAh) for energy dosage.
SYST:ENER:DOS:ENER:TYP {< type >}
Setting of energy type for Energy Dosage (Test). Parameters are 0 = Active, 1 = Reactive, 2 =
Apparent.
SYST:ENER:DOS:ENER:TYP?
Request for energy type for Energy Dosage (Test). Answers are 0 = Active, 1 = Reactive, 2 =
Apparent.
SYST:ENER:DOS:RES:SAV
Saving the results of Dial (Dosage) test. Return values are: 0
– result available, 1 – saved result, 2 –
no result available.
SYST:ENER:DOS:RES?
Request for overall error of Dial (Dosage) test after finishing this test.
SYST:ENER:DOS:DUT:ENER:K {<x>,<y>}
Setting the values of Dial Test from el. meter. Parameters are: x = value type, 0
– Initial, 1 – Final, 2 –
their difference. y = actual value of chosen type by x.
SYST:ENER:DOS:DUT:ENER:K?
Request for all values of meter dial: Initial, Final, and differential (delta).
SYST:ENER:DOS:MAXD:WIN {<window>}
Setting the value of Maximum Demand Test’s window parameter. This parameters represents setting
of Sliding or Fixed type of window.
SYST:ENER:DOS:MAXD:PER {<period>}
Setting the value of Maximum Demand Test’s period parameter. This parameters represents time
stated in minutes.
SYST:ENER:DOS:MAXD:SUBP {<subperiod>}
Setting the value of Maximum Demand Test’s sub-period parameter.
SYST:ENER:DOS:DUT:MAXD:K {<register>}
Setting the register value of Maximum Demand Test for DUT (Device Under Test).
SYST:ENER:DOS:RES:MAXD?
Request for error of Maximum Demand Test after finishing this test.
SYST:ENER:DOS:STAT:MAXD:K?
Request for state of Maximum Demand Test.
SYST:ENER:DOS:STAT:MAXD:K:TEMP?
Request for the recently integrated value of power (Maximum Demand Test).
SYST:ENER:DOS:STAT:MAXD:WIN:BEG?
Request for the beginning timestamp of Maximum Demand Test being in progress.
SYST:ENER:DOS:STAT:MAXD:WIN:END?
Request for the ending timestamp of Maximum Demand Test being in progress.