Parker Balston Model 9150 

 

Hydrogen Generator

General Description

THIS PRODUCT IS SHELF LIFE SENSITIVE.  DO NOT STORE FOR MORE THAN 3 MONTHS 
PRIOR TO INSTALLATION AND OPERATION.  These instructions must be thoroughly 
read and understood before installing and operating this product.  Any modification of 
the product will void the warranty.  Failure to operate this product in accordance with the 
instructions set forth in this manual could jeopardize the safety of the operator.

If you have any questions or concerns, please call Technical Services Department at      
800-343-4048, 8AM to 5PM Eastern Time (North America only).  For other locations, please 
contact your local representative.  Send e-mail to: [email protected].

Do not use a hydrogen gas storage vessel in conjunction with the hydrogen generator. 
Stored quantities of hydrogen pose an explosion hazard.

Normal precautions for any hydrogen supply should be taken when using the hydrogen 
generator. DO NOT USE IN A SEALED OR UNVENTED ROOM.

CAUTION: DO NOT USE AN OPEN FLAME OR OTHER IGNITION SOURCE WITHIN               
10 INCHES (25cm) OF THE OXYGEN VENT!

The Hydrogen Generator Model A915000JA-L1555 (see Figure 1), is a completely engineered 
system designed to produce ultra-pure, dry hydrogen gas from deionized water using a Pro-
ton Exchange Membrane Cell.  The hydrogen generator utilizes the principle of electrolytic 
dissociation of water and hydrogen proton conduction through the membrane.  The hydrogen 
stream is dried using a coalescing filter, a drying tube, and desiccant.  The outlet pressure of 
the unit is 90 psig (6.2 barg) with a maximum hydrogen delivery flow rate of 160 cc/min for 
Model 9150.

The hydrogen generators are ideal for use as fuel gas in Flame Ionization Detectors (FIDs), 
Trace Hydrocarbon Analyzers, and other general laboratory equipment requiring a source of 
pure hydrogen.

The 9150 Hydrogen Generator is certified to the electrical safety requirements as specified 
by the IEC, CSA, and UL standards. These units bear both the CSA and UL markings on the 
product label. Product supplied to Europe carries the CE mark   (230 VAC units only).  The 

Figure 1 - Hydrogen Generator with 

flow bar graph and water check

Regulatory 
Compliance

Parker Balston  

A915000JA-LI555 

Hydrogen Generator

Installation, Operation, and Maintenance Manual

Technical 

Information

Bulletin TI-A915000JA-L1555

Parker Hannifin Corporation

Industrial Gas Filtration and Generation  

Division

Lancaster, NY  • 1-800-343-4048 

Summary of Contents for A915000JA-LI555

Page 1: ...ITION SOURCE WITHIN 10 INCHES 25cm OF THE OXYGEN VENT The Hydrogen Generator Model A915000JA L1555 see Figure 1 is a completely engineered system designed to produce ultra pure dry hydrogen gas from d...

Page 2: ...on the back of the unit Desiccant Wipe any drops of moisture with a lint free cloth or paper towel from the desic cant cap and housing Remove the desiccant cartridge from the sealed bag Use a drop of...

Page 3: ...rator Pressure Regulator The pressure is controlled using the Pressure Adjust control knob located on the front panel exposed when opening the front gray panel door see Figure 7 The pressure can be mo...

Page 4: ...move all of the red tape that was applied to the unit for shipping purposes Do not let the deionizer bags dry out Open the gray panel door on the front of the generator to expose the generator control...

Page 5: ...t a Mass Leak is present please follow the procedure below 1 Close the Shutoff Valve see Figure 7 shown in the OPEN position 2 Turn Power switch located in the rear of the generator in the power recep...

Page 6: ...r further instructions The generator is designed to be run continuously 24 hours 7 days per week If the unit is to be idle for more than 24 hours the generator should be turned off To store the unit 1...

Page 7: ...is may be caused either from a leak in the piping to the equipment or it may indicate the volume that is being filled is too large for the generator The High Hydrogen Flow Indicator will stay illumina...

Page 8: ...corresponds to the changes in the current going through the water Low Water Indicator If the water level drops too low in the water reservoir the generator will stop producing hydrogen Blue or orange...

Page 9: ...voirs are filled to the Full level Water must have a minimum resistivity of 2 MegOhm cm If the Low Water indicator see Figure 7 illuminates the generator will stop producing hydro gen to safeguard the...

Page 10: ...is in the OPEN position see Figure 8 in order to vent the hydrogen from the generator 2 Lift the top cover and unscrew the water reservoir caps Lift the deionizer bag out of the reservoir along with...

Page 11: ...Temperature 40 100 F 4 38 C Electrical Requirements 2 100 120 200 240 VAC 47 63 Hz Power Consumption 480 Watts 110 VAC 460 Watts 220 VAC Fuse 110 VAC T 4 Amp 125 V Fuse 220 VAC T 2 Amp 250 V Product D...

Page 12: ...eck external piping for leaks Hydrogen demand exceeds generator capacity Check capacity of equipment and piping downstream and install a flow control device Check internal system for piping leaks Chec...

Page 13: ...Hydrogen Generator Symbol Description POWER INDICATOR HYDROGEN FLOW INDICATOR NO HYDROGEN FLOW INDICATOR HIGH HYDROGEN FLOW MASS LEAK or EXCESSIVE FLOW RATE INDICATOR CHANGE WATER INDICATOR LOW WATER...

Page 14: ...ns to the skin found on inside of some units Under certain conditions metal cabinet will be warm to touch 70 C 3 Explanation of Warning Symbols Parker Manufacturing UK Ltd Gas Separation Filtration Di...

Reviews: