Parker Balston A915000JA-LI555 Installation, Operation And Maintenance Manual Download Page 5

Parker Hannifin Corporation

Industrial Gas Filtration and Generation  Division

Lancaster, NY  • 1-800-343-4048 

www.parker.com/igfg

5

    

Bulletin TI-A915000JA-L1555 

Parker Balston Model 9150 

 

Hydrogen Generator

Mass Leak

lock the dial.

4

  Turn the 

Power Switch

 ON (located in the rear of the generator, in the power receptacle).

Note

:  The generator is now producing hydrogen at the requested pressure.  As the unit approaches the 

desired pressure, the 

Hydrogen Flow Indicator

 light may flicker, dim or even go out. This is normal oper-

ation.  It will take approximately 15 minutes to reach full operating pressure internally.

Connect the external piping to the generator.  While monitoring the 

Output Pressure Gauge

 on 

the front panel, slowly turn the 

Shutoff Valve

 to the OPEN position (see Figure 7). 

6

  The pressure will drop initially, the magnitude of which is dependent upon the length and diame-

ter of tubing used, and volume of the connected equipment downstream.  After a short time the 
pressure should begin to recover.  Once pressure has recovered, open the outlet valve a little 

Follow the instructions listed in the 

Initial Startup

 section.  If the pressure continues to drop, the 

High 

Hydrogen Flow

 (Mass Leak) light will illuminate (see Figure 7).  A Mass Leak may be caused either 

from a leak in the piping to the equipment or it may indicate that the volume being filled is too large 
for the generator.  If it is determined that a Mass Leak is present, please follow the procedure below:

1

  Close the 

Shutoff Valve 

(see Figure 7, shown in the OPEN position).

2

  Turn Power switch (located in the rear of the generator in the power receptacle) OFF and then 

turn Power switch ON.

3

  Re-pressurize the generator. 

4

  Repeat the Initial Startup procedure three times to verify the leak.

If the system continues to detect a leak, check all the external tubing connections and make sure that 
the volume is not too large for the generator purchased.  If the leak persists, check the fittings inside 
the generator, especially the o-ring seal on the desiccant cap.  Make sure the desiccant cap is tight.

If there are no leaks found in internal or external connections, install a flow control valve or needle 
valve downstream from the generator, as close to the outlet as possible.  Set the flow control for the 
desired flow rate making sure not to exceed the maximum flow rate of the generator.

Caution:

  

Do not use commercial leak detection fluids on any Lexan

®

 plastic parts in the generator.

  

These fluids may cause cracking and crazing which results in leaks.  Use a solution of 4-5 ml of liquid 
dish soap in one liter of water.

Caution

Use extreme care when checking for internal leaks.

 Due to the nature of the gases 

involved use a protective face shield or other appropriate equipment whenever servicing or trouble-
shooting the generator with the cover removed. 

Figure 9 - To remove the Drain Port Insert, press down on the 

metal clip to release the insert, and pull.

Figure 8 - Insert the Drain Port Insert 

along with a length of tubing into the 

Drain Port in the Front

Summary of Contents for A915000JA-LI555

Page 1: ...ITION SOURCE WITHIN 10 INCHES 25cm OF THE OXYGEN VENT The Hydrogen Generator Model A915000JA L1555 see Figure 1 is a completely engineered system designed to produce ultra pure dry hydrogen gas from d...

Page 2: ...on the back of the unit Desiccant Wipe any drops of moisture with a lint free cloth or paper towel from the desic cant cap and housing Remove the desiccant cartridge from the sealed bag Use a drop of...

Page 3: ...rator Pressure Regulator The pressure is controlled using the Pressure Adjust control knob located on the front panel exposed when opening the front gray panel door see Figure 7 The pressure can be mo...

Page 4: ...move all of the red tape that was applied to the unit for shipping purposes Do not let the deionizer bags dry out Open the gray panel door on the front of the generator to expose the generator control...

Page 5: ...t a Mass Leak is present please follow the procedure below 1 Close the Shutoff Valve see Figure 7 shown in the OPEN position 2 Turn Power switch located in the rear of the generator in the power recep...

Page 6: ...r further instructions The generator is designed to be run continuously 24 hours 7 days per week If the unit is to be idle for more than 24 hours the generator should be turned off To store the unit 1...

Page 7: ...is may be caused either from a leak in the piping to the equipment or it may indicate the volume that is being filled is too large for the generator The High Hydrogen Flow Indicator will stay illumina...

Page 8: ...corresponds to the changes in the current going through the water Low Water Indicator If the water level drops too low in the water reservoir the generator will stop producing hydrogen Blue or orange...

Page 9: ...voirs are filled to the Full level Water must have a minimum resistivity of 2 MegOhm cm If the Low Water indicator see Figure 7 illuminates the generator will stop producing hydro gen to safeguard the...

Page 10: ...is in the OPEN position see Figure 8 in order to vent the hydrogen from the generator 2 Lift the top cover and unscrew the water reservoir caps Lift the deionizer bag out of the reservoir along with...

Page 11: ...Temperature 40 100 F 4 38 C Electrical Requirements 2 100 120 200 240 VAC 47 63 Hz Power Consumption 480 Watts 110 VAC 460 Watts 220 VAC Fuse 110 VAC T 4 Amp 125 V Fuse 220 VAC T 2 Amp 250 V Product D...

Page 12: ...eck external piping for leaks Hydrogen demand exceeds generator capacity Check capacity of equipment and piping downstream and install a flow control device Check internal system for piping leaks Chec...

Page 13: ...Hydrogen Generator Symbol Description POWER INDICATOR HYDROGEN FLOW INDICATOR NO HYDROGEN FLOW INDICATOR HIGH HYDROGEN FLOW MASS LEAK or EXCESSIVE FLOW RATE INDICATOR CHANGE WATER INDICATOR LOW WATER...

Page 14: ...ns to the skin found on inside of some units Under certain conditions metal cabinet will be warm to touch 70 C 3 Explanation of Warning Symbols Parker Manufacturing UK Ltd Gas Separation Filtration Di...

Reviews: