8

 

Hydrogen Purity

Change Water Indicator

 - If the water quality in the reservoir reaches the Low level, the 

Change Water Indicator (see Figure 7) will illuminate.  The water in the unit must be changed 
and the deionizer bags must be replaced. 

Water Check Bar Graph 

- If the Mode Button is pressed in on the Parker Balston 9150, the 

Bar Graph will display the Water Check quality (see Figure 10). The Bar Graph indicates the 
water quality level from High (>2 MegOhm/cm) to Low. This quality level corresponds to the 
changes in the current going through the water.

Low Water Indicator

 - If the water level drops too low in the water reservoir, the generator will 

stop producing hydrogen.

Blue (or orange) colored indicator beads are located in the bottom half of the desiccant car-
tridge.  The blue beads turn gray or beige (the orange beads turn clear) as water is adsorbed 
in the cartridge.  When this happens, the desiccant should be changed immediately, or the 
water content in the hydrogen stream will increase dramatically. 

Note

:  If low moisture (1 ppm or less) is required, the desiccant cartridge must be replaced be-

fore all the indicator beads change color.  An external water trap can be used downstream from 
the generator for increased water removal.

Water Quality 

Indicators

Figure 11: Diagnostic indicators

BAR GRAPH

MODE Button

POWER Indicator

HYDROGEN FLOW

NO HYDROGEN FLOW

s

s

s

s

s

Summary of Contents for A915000JA-LI555

Page 1: ...ITION SOURCE WITHIN 10 INCHES 25cm OF THE OXYGEN VENT The Hydrogen Generator Model A915000JA L1555 see Figure 1 is a completely engineered system designed to produce ultra pure dry hydrogen gas from d...

Page 2: ...on the back of the unit Desiccant Wipe any drops of moisture with a lint free cloth or paper towel from the desic cant cap and housing Remove the desiccant cartridge from the sealed bag Use a drop of...

Page 3: ...rator Pressure Regulator The pressure is controlled using the Pressure Adjust control knob located on the front panel exposed when opening the front gray panel door see Figure 7 The pressure can be mo...

Page 4: ...move all of the red tape that was applied to the unit for shipping purposes Do not let the deionizer bags dry out Open the gray panel door on the front of the generator to expose the generator control...

Page 5: ...t a Mass Leak is present please follow the procedure below 1 Close the Shutoff Valve see Figure 7 shown in the OPEN position 2 Turn Power switch located in the rear of the generator in the power recep...

Page 6: ...r further instructions The generator is designed to be run continuously 24 hours 7 days per week If the unit is to be idle for more than 24 hours the generator should be turned off To store the unit 1...

Page 7: ...is may be caused either from a leak in the piping to the equipment or it may indicate the volume that is being filled is too large for the generator The High Hydrogen Flow Indicator will stay illumina...

Page 8: ...corresponds to the changes in the current going through the water Low Water Indicator If the water level drops too low in the water reservoir the generator will stop producing hydrogen Blue or orange...

Page 9: ...voirs are filled to the Full level Water must have a minimum resistivity of 2 MegOhm cm If the Low Water indicator see Figure 7 illuminates the generator will stop producing hydro gen to safeguard the...

Page 10: ...is in the OPEN position see Figure 8 in order to vent the hydrogen from the generator 2 Lift the top cover and unscrew the water reservoir caps Lift the deionizer bag out of the reservoir along with...

Page 11: ...Temperature 40 100 F 4 38 C Electrical Requirements 2 100 120 200 240 VAC 47 63 Hz Power Consumption 480 Watts 110 VAC 460 Watts 220 VAC Fuse 110 VAC T 4 Amp 125 V Fuse 220 VAC T 2 Amp 250 V Product D...

Page 12: ...eck external piping for leaks Hydrogen demand exceeds generator capacity Check capacity of equipment and piping downstream and install a flow control device Check internal system for piping leaks Chec...

Page 13: ...Hydrogen Generator Symbol Description POWER INDICATOR HYDROGEN FLOW INDICATOR NO HYDROGEN FLOW INDICATOR HIGH HYDROGEN FLOW MASS LEAK or EXCESSIVE FLOW RATE INDICATOR CHANGE WATER INDICATOR LOW WATER...

Page 14: ...ns to the skin found on inside of some units Under certain conditions metal cabinet will be warm to touch 70 C 3 Explanation of Warning Symbols Parker Manufacturing UK Ltd Gas Separation Filtration Di...

Reviews: