Using the Script Menu
MT8850A / MT8852A Bluetooth Test Set
13000-000109
8-9
January 2003
Setting Up General Script Parameters
5.
Repeat this procedure until all of the required offsets have been specified.
The values entered in the table are used to interpolate the offsets for the intermediate
channels that were not actually specified. It should also be noted that the offsets are used
to perform compensations for either the receiver tests in the script or for the sensitivity
tests (single sensitivity), but never for both. For the receiver tests, the output power
measurements are compensated, and for the sensitivity test the transmitter is
compensated.
Summary of Contents for MT8850A
Page 2: ......
Page 6: ...iv 13000 000109 January 2003 Declaration of Conformity...
Page 7: ...13000 000109 v January 2003...
Page 23: ...Installation MT8850A MT8852A Bluetooth Test Set 13000 000109 3 5 January 2003 Rack Mounting...
Page 127: ...Top line Messages MT8850A MT8852A Bluetooth Test Set 13000 000109 A 9 January 2003...
Page 128: ......
Page 130: ...MT8850A MT8852A Bluetooth Test Set Preset Values B 2 13000 000109 January 2003...
Page 148: ...MT8850A MT8852A Bluetooth Test Set Connector Care and Handling D 4 13000 000109 January 2003...
Page 153: ......