MT8850A / MT8852A Bluetooth Test Set
Configuring the MT8850A/MT8852A
7-14
13000-000109
Setting up the MT8850A/MT8852A
January 2003
Performing Continuous Wave Measurement
The MT8850A/MT8852A can measure a fixed frequency signal to calibrate power,
frequency, and modulation. The CW measurement mode is intended only for the
measurement of continuous non-packetized signals and does not support triggering.
5. Press
the
hard key.
6. Choose
“MT8852A”.
7.
Move the cursor to “CW Measurement” and press the
key.
8.
CW Measurement is commenced using the [
Enable
] soft key and switched off using
the [
Exit
] soft key or pressing the
button.
“Gate width” defines the measurement window. Settings can be made between 1 and the
default setting of 10 ms.
“Channel frequency” may be replaced with “Channel number” depending on the setting
that has been made at “Display frequency as” on the All scripts setup page (detailed in the
following section of this manual.) The default channel number is 2, with settings possible
in the range of –2 to 98.
Note:
No results will be displayed in the modulation field if the received power level
drops below -40 dBm.
Summary of Contents for MT8850A
Page 2: ......
Page 6: ...iv 13000 000109 January 2003 Declaration of Conformity...
Page 7: ...13000 000109 v January 2003...
Page 23: ...Installation MT8850A MT8852A Bluetooth Test Set 13000 000109 3 5 January 2003 Rack Mounting...
Page 127: ...Top line Messages MT8850A MT8852A Bluetooth Test Set 13000 000109 A 9 January 2003...
Page 128: ......
Page 130: ...MT8850A MT8852A Bluetooth Test Set Preset Values B 2 13000 000109 January 2003...
Page 148: ...MT8850A MT8852A Bluetooth Test Set Connector Care and Handling D 4 13000 000109 January 2003...
Page 153: ......