MT8850A / MT8852A Bluetooth Test Set
Connector Care and Handling
D-2
13000-000109
January 2003
DUT Connector
Type
Anritsu Gauging
Set Model
Pin Depth
(inches)
Pin Depth
Gauge Reading
N - Male
N - Female
01-163
207 -0.000
+0.030
207 +0.000
-0.030
WSMA - Male
WSMA - Female
3.5 mm - Male
3.5 mm - Female
01-162
-0.000
-0.010
Same as pin
depth
Allowable DUT connector pin depth
Avoid over torquing connectors
Over torquing connectors is destructive; it may damage the connector center pin. Finger-
tight is usually sufficient for Type N connectors.
Always
use a connector torque wrench (8
inch-pounds) when tightening WSMA, K or V type connectors.
Never use pliers to tighten
connectors.
Avoid mechanical shock
Precision connectors are designed to withstand years of normal bench handling. However,
do not drop or otherwise treat them roughly. Mechanical shock will significantly reduce
their service life.
Do not disturb Teflon tuning washers on connector center pins
The center conductor on many RF component connectors contains a small Teflon washer
that is located near the point of mating. This washer compensates for minor impedance
discontinuities at the interface.
Do not disturb this washer.
The location of the washer is
critical to the performance of the RF component.
Cleaning connectors
The precise geometry that makes possible the RF component's high performance can
easily be disturbed by dirt and other contamination adhering to the connector interfaces.
When not in use, keep the connectors covered.
To clean the connector interfaces, use a clean cotton swab that has been dampened with
denatured alcohol.
NOTE:
Most cotton swabs are too large to fit in the smaller
connector types. In these cases, it is necessary to peel off most
of the cotton and then twist the remaining cotton tight. Be sure
that the remaining cotton does not get stuck in the connector.
Cotton swabs of the appropriate size can be purchased through
a medical laboratory-type supply center.
The following are some important tips on cleaning connectors:
•
Use only denatured alcohol as a cleaning solvent
•
Do not use excessive amounts of alcohol as prolonged drying of the connector may be
required
•
Never put lateral pressure on the center pin of the connector
•
If installed, do not disturb the Teflon washer on the center conductor pin
Summary of Contents for MT8850A
Page 2: ......
Page 6: ...iv 13000 000109 January 2003 Declaration of Conformity...
Page 7: ...13000 000109 v January 2003...
Page 23: ...Installation MT8850A MT8852A Bluetooth Test Set 13000 000109 3 5 January 2003 Rack Mounting...
Page 127: ...Top line Messages MT8850A MT8852A Bluetooth Test Set 13000 000109 A 9 January 2003...
Page 128: ......
Page 130: ...MT8850A MT8852A Bluetooth Test Set Preset Values B 2 13000 000109 January 2003...
Page 148: ...MT8850A MT8852A Bluetooth Test Set Connector Care and Handling D 4 13000 000109 January 2003...
Page 153: ......