Configuring the MT8850A/MT8852A
MT8850A / MT8852A Bluetooth Test Set
13000-000109
7-15
January 2003
Setting up the MT8850A/MT8852A
Setting the Range Hold
The Range Hold facility allows you to fix the MT8850A/MT8852A measuring receiver on a
given range and disable the auto-range feature.
1. Press
the
hard key.
2.
Choose “MT8850A/ MT8852A”.
3. Press
the
key.
The MT8850A/MT8852A Configuration screen is displayed.
4.
Choose “RX/TX settings”.
5. Press
the
key.
The MT8850A/MT8852A RX/TX settings screen is displayed.
6.
Highlight the Range hold value.
The values are Auto, 1 2, 3, 4, 5 and 6.
•
Auto is the default setting.
•
RGH1 represents a range of +22 to +7 dBm.
•
RGH 2 represents a range of +9 to -3 dBm.
•
RGH 3 represents a range of +5 to -7 dBm.
•
RGH 4 represents a range of -4 to -16 dBm.
•
RGH 5 represents a range of -12 to -26 dBm.
•
RGH 6 represents a range of -24 to -35 dBm.
When any of the numbered ranges is displayed, the value will be displayed on the top
line of the screen.
When the value is Auto, the
[Range hold]
soft key is displayed and will cause the
current range to be held when pressed.
7.
Repeatedly press the
key until the required value is displayed.
Summary of Contents for MT8850A
Page 2: ......
Page 6: ...iv 13000 000109 January 2003 Declaration of Conformity...
Page 7: ...13000 000109 v January 2003...
Page 23: ...Installation MT8850A MT8852A Bluetooth Test Set 13000 000109 3 5 January 2003 Rack Mounting...
Page 127: ...Top line Messages MT8850A MT8852A Bluetooth Test Set 13000 000109 A 9 January 2003...
Page 128: ......
Page 130: ...MT8850A MT8852A Bluetooth Test Set Preset Values B 2 13000 000109 January 2003...
Page 148: ...MT8850A MT8852A Bluetooth Test Set Connector Care and Handling D 4 13000 000109 January 2003...
Page 153: ......