Serial Communications
|
4-7
Function: 0D Ionizer Parameters
•
Sets the ionizer parameters for the head.
Parameters
Emis ElEnergy Focus Repeller
Emis
ElEnergy
Focus
Repeller
Range:
Emission current. Range: 1.0E-04 to 1.0E-02.
Electron energy. Range: -150.0 to -30.0
Focus voltage. Range: -200.0 to +10.0.
Repeller or CIS lens voltage (if connected to a CIS head).
-200.0 to +10.0.
Returns
If no parameter supplied, then the ionizer parameter string is returned.
Function: 0E
Autozero
•
Turn the autozero option on or off.
When turned on, the scanning time will double. For each data point, two readings are taken.
One normal, and one with the mass filter blocking all ions.
Parameters
1
0
(Turn on)
(Turn off)
Returns
If no parameter supplied, returns current value.
Function: 0F
Set Time
•
Set the real time clock.
Time and date are not maintained when powered off.
Parameters
month/day/year hour:minute:second
month
day
year
hour
minute
second
1 to 12
1 to 31
four digit year, e.g. 1998
00 to 23 (This must be in 24 hour format)
00 to 59
00 to 59
Returns
If no parameter supplied, returns current date and time in the format:
month/day/year hour:minute:second.hundredths
Summary of Contents for Dycor Dymaxion Mass Spectrometer
Page 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Page 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Page 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Page 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Page 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Page 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Page 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...