I/O Board Installation
|
A-11
I/O Option Board Specifications
Digital Outputs
Number of channels:
6
Dry contact relays
Maximum current:
500 mA
Maximum voltage:
30 VAC or 60 VDC
Digital Inputs
Number of channels:
4
Isolation:
Optical, 5000V RMS
Input “ON” range:
2 - 30VDC or AC
Input “OFF” range:
< 1VDC or AC
Analog Inputs
Number of channels:
2
Resolution:
12 bits (4096 steps)
Input Range:
Current (0-20 mA) or Voltage (0-10 VDC)
Jumper-selectable
Isolation:
Single-Ended
Voltage input range:
0 - 10 VDC
Minimum undervoltage:
-15 V
Maximum overvoltage:
15 V
Analog Outputs
Number of channels:
2
Resolution:
12 bits (4096 steps)
Output ranges:
Current (4-20 mA) or Voltage (0 - 5 VDC, 0 - 10
VDC)
Jumper selectable
Isolation:
Optical, 500 VDC
Current output range:
4 - 20 mA
Maximum load:
600 ohms
Voltage output range:
0 - 5 VDC
Minimum load:
2K ohms
Voltage output range:
0 - 10 VDC
Minimum load:
2K ohms
Summary of Contents for Dycor Dymaxion Mass Spectrometer
Page 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Page 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Page 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Page 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Page 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Page 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Page 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...