Installation
|
3-3
Connect Sensor to the Vacuum System
Open Source and AMETEK Vacuum Chamber
1.
Using a new copper gasket, place the sensor into the AMETEK vacuum
chamber so that the source resides inside the long end of the chamber.
Be sure the repeller clears the chamber walls.
2.
Orient the barrel notch in the 12 o’clock position (Figure 3-1).
3.
Bolt the source flange in place using the (6) ¼-28 x 1” screws and
washer provided. Tighten in a star pattern to ensure a good seal.
4.
Install the appropriate inlet components to the source end of the vac-
uum chamber if they are not already installed. These consist of a ¾”
thick VCR flange, a blank flange, a manual high-conductance valve or
electro-pneumatic (EP) high-conductance valve, two copper gaskets, and
fastener hardware. Inlet details
are found in the Chapter 5.
Figure 3-1.
Source pins. View of
flange (atmosphere
side).
Analyzer Head
Mass Filter
AMETEK
Vacuum Manifold
Copper Gasket
(6) 1/4-28 Screws
Barrel Notch
Up
Analyzer Head
Mass Filter
AMETEK
Vacuum Manifold
Copper Gasket
(6) 1/4-28 Screws
Barrel Notch
Up
Figure 3-2.
Connecting the open-
source to the AMETEK
vacuum chamber.
Summary of Contents for Dycor Dymaxion Mass Spectrometer
Page 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Page 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Page 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Page 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Page 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Page 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Page 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...