4-12 | Dycor Dymaxion Mass Spectrometer
Operation
In order to send commands to the Dymaxion, the host must log in by sending the
00L function.
The Dymaxion is organized into 32 data channels. Each of the channels can be
set to either tabular (a single mass) or to analog, bar, total pressure, analog input,
digital input, or disabled. A “scan” consists of acquiring data for all active chan-
nels. The date and time are stored with each scan.
All parameters are retained when the Dymaxion is turned off. The time and date
are
not
updated while the power is off.