Maintenance & Troubleshooting
|
6-5
Removing the old filament
1.
Remove the six (6) ¼-28 bolts that fasten the quadrupole head 10-pin
feedthrough to the vacuum system.
2.
Remove and discard the copper gasket.
DO NOT TRY TO REUSE THE GASKET.
3.
Position the quadrupole head so that it is pointing up and is secure.
4.
The ion volume on the enclosed source must be removed to replace
the filament. Using a hex wrench, remove the lock screw on the col-
lar holding the ion volume to the source (see Figure 6-2). Carefully
remove the ion volume and set it asi
de.
Filament Hold-Down
Screw
Filament Hold-Down
Screw
Ion Volume
Filament Plate Tab
Lock Ring
Ion Volume
Lock Screw
Figure 6-2. Dymaxion ion volume and filament.
5.
The filament plate is held in place by two (2) stainless steel rods that
carry the current to the filament. Using the hex wrench provided, loos-
en the two screws enough to rotate the filament plate counterclock-
wise and remove it by sliding the plate over the screw heads (see
Figure 6-2). Be sure to remove all residual pieces of the old filament.
Summary of Contents for Dycor Dymaxion Mass Spectrometer
Page 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Page 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Page 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Page 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Page 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Page 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Page 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...