Serial Communications
|
4-3
Function: 00
Status
This is the only command that will work before you log in. Either 00R or 00L is required after
power-up to log in.
If no parameter is supplied, it returns the current Dymaxion status as a two two-digit hex values.
For example, if you send “>010060[CR]” the response may be “A00 0040[CR].”
Status Byte #1
Bit 0:
Bit 1:
Bit 2:
Bit 3:
Bit 4:
Bit 5:
Bit 6:
Bit 7:
0x01
0x02
0x04
0x08
0x10
0x20
0x40
0x80
Filament shorted
Filament open
Filament half open
Filament off because of over-pressure
RF Amplifier tune failure
Interlock not set
Over temperature failure
General failure
Status Byte #2
Bit 0:
Bit 1:
Bit 2:
Bit 3:
Bit 4:
Bit 5:
Bit 6:
Bit 7:
0x01
0x02
0x04
0x08
0x10
0x20
0x40
0x80
+15 volt supply failure
-15 volt supply failure
+80 volt supply failure
+200 volt supply failure
-200 volt supply failure
Reserved
Reserved
Reserved
If sending a subfunction code in the first byte of the data block:
Subfunction: R
Reset
Logs into the Dymaxion. Stops scanning, resets system and clears errors. All scan data, chan-
nel settings, and tune parameters are cleared.
Subfunction: L
Log In
Logs into the Dymaxion. Stops scanning, resets system and clears errors.
Subfunction: O
Log Out
Logs out from the Dymaxion. Stop scanning, turn off filament, multiplier, and degas mode.
Subfunction: I
Information
Gets information string.
Returns
“Dycor Dymaxion 1.0” Where 1.0 is the firmware version.
Subfunction: A
Repeat last ACK
Repeats the last acknowledge message sent. If an ACK message is received with errors, use
this to retransmit.
Returns
A copy of the last ACK message sent.
Summary of Contents for Dycor Dymaxion Mass Spectrometer
Page 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Page 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Page 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Page 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Page 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Page 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Page 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Page 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...