Configuration and Calibration
55
4.5 Configuration Menu (Continued)
Function Item
(Display)
Description & comment
Menu Selec-
tion Choices
Affected parameters
(changing this pa-
rameter will result
in a change these
parameters)
Fct. 1.00 Level Menu
(LEVEL)
System configuration for level.
You must start with this Level
menu to configure and calibrate
the Impulse TDR
Fct. 1.01 Level Units
(UNITS)
Selects the units of measurement
for all level and distance based
outputs, calculations and configu-
ration entries.
Changing units will produce an
automatic conversion of all level
and distance parameters.
FEET
INCHES
(Default)
METERS
CM (centi-
meters)
MM (milli-
meters)
Fct 1.02: Tank
Height
Fct 1.03: Probe
Length
Fct 1.04: Sensor
Offset
Fct 1.06: Upper Dead
Zone
Fct 1.07: Lower
Dead Zone
Fct 3.03: In Point
Fct. 1.02 Tank Height
(TANK HT)
Required for accurate Level mea-
surement (Level = Tank
Offset – Distance). Edit to reflect
the distance from tank bottom to
tank top.
If other than sensor mounting is
used for the top reference the Off-
set function must be used.
User editable
value
Default =
Probe Length
Fct. 1.03 Probe
Length
(PRB LEN)
Factory set to probe length as
ordered. Edit the menu if probe
length is changed.
On cable sensors, this includes the
weight.
User editable
value
Default =
600”
Summary of Contents for Impulse GW Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 16: ...Section 2...
Page 25: ...Installation 19 2 6 Dual Compartment Housing...
Page 26: ...Section 3...
Page 56: ...Section 4...
Page 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Page 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Page 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Page 86: ...Section 6...
Page 92: ...Section 7...
Page 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Page 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Page 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Page 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...