Configuration and Calibration
49
Tank Mapping
3.7.9 Tank Mapping
Clicking the Tank Mapping button opens the Tank Mapping screen below.
• 2 options are available Map Tank and Restore Factory Map.
• Map Tank:
• Enter Map to Distance. This is the distance from the bottom
of the mounting that will be recorded and subtracted from the
measurement profile. Do not map within 8 inches of the level
without consulting factory.
• Click Map Tank button
• Restore Factory Map:
• Click this button if an incorrect or suspected incorrect tank map
has been performed.