Configuration Parameters (Continued)
Point
Calibration
(F6)
Allows the correction of the level measurement by
entering the correct distance from the bottom of the sensor
mounting to the liquid surface. This offsets the reading as
required to produce the corrected measurement.
Consult factory before using this function.
DAC Trim
Allows a field reference meter to be connected to the
transmitter for adjusting transmitter output current
4 mA Trim:
3.5 to 4.5
20 mA Trim
19.5 to 20.5
Strapping
Table
The strapping table is a look-up table of level vs. volume
that is entered by the user based on known parameters.
Use this function for volume measurement only
User editable values
Configure
Meter
Configures the 7-Digit Display to show Level, Volume.4-20
mA, % of Range, Distance. Either as a constant display or
scrolling all enabled choices
TDR
Configuration
Configures the current output for Near Zone and Lost Echo
conditions
Lost Echo occurs when the unit cannot find a valid level or
the end of the probe and represents an error condition.
Near Zone occurs when a level is tracked into the Upper
Dead Zone. When this occurs the current will go to the
selected value and remain there until the level falls back
into the measurable range
NOTE: If the system is powered on or “signal acquire” is
activated with the level in the Near Zone, it will not be
detected and an erroneous reading may result.
Lost Echo:
Low Current (3.7
mA - Default)
High Current (22
mA)
Near Zone:
Low Current (3.7
mA)
High Current (22
mA - Default)
Restore
factory
defaults
Selecting Yes will reset all configured parameters to the
Factory Default values
YES / NO
Acquire target Initiate search mode. Can be used if unit has detected
incorrect reflection as level.
Upgrade
software
Used to upgrade onboard software to the latest version.
Only used with factory consultation
View tank
profile data
Opens “Tank Profile Data” window to display a graph of
the active signal
Summary of Contents for Impulse GW Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 16: ...Section 2...
Page 25: ...Installation 19 2 6 Dual Compartment Housing...
Page 26: ...Section 3...
Page 56: ...Section 4...
Page 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Page 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Page 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Page 86: ...Section 6...
Page 92: ...Section 7...
Page 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Page 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Page 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Page 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...