Impulse™ Series
24
3.3 Configuration Menu (Continued)
Function Item
(Display)
Description & comment
Menu Selection
Choices
Affected param-
eters
(changing this pa-
rameter will result
in a change these
parameters)
Fct. 1.00 Level
Menu
(LEVEL)
System configuration for level.
You must start with this Level
menu to configure and calibrate
the Impulse TDR
Fct. 1.01 Level
Units
(UNITS)
Selects the units of measurement
for all level and distance based
outputs, calculations and configu-
ration entries.
Changing units will produce an
automatic conversion of all level
and distance parameters.
FEET
INCHES (De-
fault)
METERS
CM (centimeters)
MM (millimeters)
Fct 1.02: Tank
Height
Fct 1.03: Probe
Length
Fct 1.04: Sensor
Offset
Fct 1.06: Upper
Dead Zone
Fct 1.07: Lower
Dead Zone
Fct 5.02: LRV
Fct 5.03: URV
Fct 3.03: In Point
Fct. 1.02 Tank
Height
(TANK
HT)
Required for accurate Level mea-
surement (Level = Tank
Offset – Distance). Edit to reflect
the distance from tank bottom to
tank top.
If other than sensor mounting
is used for the top reference the
Offset function must be used.
User editable
value
Default = Probe
Length
Fct. 1.03 Probe
Length
(PRB
LEN)
Factory set to probe length as
ordered. Edit the menu if probe
length is changed.
On cable sensors, this includes
the weight.
User editable
value
Default = 600”
Summary of Contents for Impulse GW Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 16: ...Section 2...
Page 25: ...Installation 19 2 6 Dual Compartment Housing...
Page 26: ...Section 3...
Page 56: ...Section 4...
Page 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Page 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Page 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Page 86: ...Section 6...
Page 92: ...Section 7...
Page 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Page 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Page 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Page 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...