27
DIAG
TEST
BACK
MAIN
STP/
NTP
ROTATE
DISP
STP
PRESS
MAIN
STP
TEMP
STP TEMP Display
CLEAR
DOWN
BACK/
CANCEL
Temperature: #C
SET
UP
SELECT
DIGIT
25.0000
MISC2
will display as shown at left.
STP/NPT
refers to the functions that allow
your selection of
standard
temperature and
pressure conditions or
normal
temperature
and pressure conditions. This feature is
generally useful for comparison purposes to
other devices or systems using different STP
parameters.
The
STP
menu is comprised of the
STP TEMP
and
STP PRESS
screens.
STP TEMP
allows you to select from 0C, 0F, K
or 0R. The arrow position will automatically
default to the currently stored value.
The
NTP
menu is comprised of the
NTP
TEMP
and
NTP PRESS
screens.
Once a selection has been made and recorded
using the
SET
button, a change acknowledgement
message will be displayed on screen.
Selecting
MAIN
will revert screen to the
Main display. If the
SET
selection is already
the currently stored value, a message
indicating that fact will appear.
STP PRESS
enables you to select from a
menu pressure settings. Use the UP/DOWN
or PAGE buttons to view the settings.
The arrow position will automatically default
to the currently stored value.
Once a selection has been made and recorded using the
SET
button, a change
acknowledgement message will be displayed on screen.
Pressing
SET
again will revert screen to the Main display. If the
SET
selection is
already the currently stored value, a message indicating that fact will appear.
STP PRESS Display
101355.0
DOWN
BACK/
CANCEL
SET
UP
SELECT
DIGIT
Pressure: PaA
CLEAR