73
Technical Data for MCP Moderate Flow Mass Flow Controllers
0 to 50 slpm Full Scale through 0 to 250 slpm Full Scale
Alicat MCP
mass flow controllers are fitted with a high performance valve for low pressure applications.
The following specifications are applicable to Alicat MCP-Series Mass Flow Controllers only.
Please Note Maximum Pressure of 80 psig.
Standard Operating Specifications (Contact Alicat for available options)
Performance
MCP Mass Flow Controller
Accuracy at calibration conditions
after tare
± (0.8% of R 0.2% of Full Scale)
High Accuracy at calibration conditions
after tare
± (0.4% of R 0.2% of Full Scale)
Repeatability
± 0.2% Full Scale
Zero Shift and Span Shift
0.02% Full Scale / ºCelsius / Atm
Operating Range / Turndown Ratio
0.5% to 100% Full Scale / 200:1 Turndown
Maximum Controllable Flow Rate
102.4% Full Scale
Typical Response Time
100 ms (Adjustable)
Warm-up Time
< 1 Second
Communications / Power
MCP Mass Flow Controller
Monochrome LCD or Color TFT
Display with integrated touchpad
Simultaneously displays Mass Flow, Volumetric Flow, Pressure and Temperature
Digital Input/Output Signal
1
Options
RS-232 Serial / RS-485 Serial / Modbus / PROFIBUS
3
Analog Input/Output Signal
2
Options
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Optional Secondary Analog
Input/Output Signal
2
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Electrical Connection Options
8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking
Supply Voltage
12 to 30 Vdc (15-30 Vdc for 4-20 mA outputs)
Supply Current
0.250 Amp
1. The
Digital Output Signal
communicates Mass Flow, Volumetric Flow, Pressure and Temperature
2. The
Analog Output Signal
and
Optional Secondary Analog Output Signal
communicate your choice of Mass
Flow, Volumetric Flow, Pressure or Temperature
3. If selecting PROFIBUS, no analog signal is available. See PROFIBUS specifications for supply voltages and
currents
(
www.alicat.com/profibus
). PROFIBUS units do not have the display.
Operating Conditions
MCP Mass Flow Controller
Mass Reference Conditions (STP)
25ºC & 14.696 psia (standard — others available on request)
Operating Temperature
−10 to +50 ºCelsius
Humidity Range
(Non–Condensing)
0 to 100%
Max. Internal Pressure (Static)
80 psig
Proof Pressure
175 psig
Mounting Attitude Sensitivity
None
Valve Type
Normally Closed
Ingress Protection
IP40
Wetted Materials
303 & 302 Stainless Steel, Viton®, Heat Cured Silicone Rubber, Glass Reinforced
Polyphenylene Sulfide, Heat Cured Epoxy, Aluminum, Gold, Brass, 430FR Stainless
Steel, Silicon, Glass.
If your application demands a different material, please contact Alicat.
Range Specific Specifications
Full Scale Flow Mass
Controller
Pressure Drop
1
at FS Flow
(psid) venting to atmosphere
Mechanical
Dimensions
Process Connections
2
MCP
50 slpm
7
4.4”H x 5.4”W x 1.6”D
1/4” NPT Female
MCP
100 slpm
20
MCP
250 slpm
60
5.0”H x 6.3”W x 1.6”D
1/2” NPT Female
1. Lower Pressure Drops Available, please see our
WHISPER-Series
mass flow controllers at
www.alicat.com/whisper
.
2. Compatible with Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and SAE
connections upon request.