28
DIAG TEST:
This diagnostic screen displays
the current internal register values, which
is useful for noting factory settings prior
to making any changes. It is also helpful
for troubleshooting with Alicat customer
service personnel.
Select the
DIAG TEST
button from the
MISC2
screen to view a list of select register values.
Pressing the
SCROLL
button will cycle the
display through the register screens. An
example screen is shown at left.
Press
ROTATE DISP
and SET to
Inverted 180
° if your device is inverted. The
display and buttons will rotate together.
BACK
MAIN
SCROLL
R8: AP Sig
7871
R9: Temp Sig
39071
R10: DP Side
9986
R11: DP Brdg
36673
R13: AP Brdg
36673
R16: Meter Func
199
R18: Power Up
32768
DEVICE UNITS
Press
DEVICE UNITS
to access menus of
units of measure for each parameter (and
totalizer if so equipped).
Scroll to the desired unit and press select.
Once selected, you will see the message
shown below. Verify that all connected
devices expect the change.
See pages 49 and 50 for a full list of
available units.
DOWN
DONE
SELECT
Mass Flow
Volumetric Flow
Pressure
Temperature
Mass Totalizer
Totalizer Time
UP
DOWN
CANCEL
SET
PAGE
SCCM
Scm3/h
Sm3/h
Sm3/d
Sin3/m
SCFH
NmL/s
UP
PRESSING SET WILL
AFFECT DISPLAY
AND SERIAL VALUES
VERIFY CONNECTED
SERIAL DEVICES
EXPECT THE CHANGE
SET
CANCEL