“Inputs/Outputs“ connector wiring
1/3
GB 00740 - Edition 09 - Januar
y 04
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600 M/MT
15
1
30
16
17
18
19
44
31
42
40
39 38
37
36
35 34
33 32
-
+
~
=
5 to 30 V
Wiring supplied
by customer
fig 1
-
+
~
=
-
+
~
=
-
+
~
=
-
+
~
=
In the case of a
local mode operation
of the
controller, the pump will run
only if
the terminal plug
(delivered with the controller)
is connected on the
“Inputs/Outputs” connector
.
The control by voltage
The control by dry
contacts with
“input control” mode
The” inputs/outputs” connector is located at the rear of the
controller (DB 44 contacts, female connector).
The inputs are considered to be activated if the AC or DC
voltage applied is between 5 and 30 Volts (fig 1).
15
1
30
28 27
25 24
16
17
18
44
41
38
37
36
35 34
32
Wiring supplied
by customer
fig 2
Start
Mode input control
Stand-by
Inhibit
Ex. fault
Stop
39
40
42
31
These inputs can be controlled by customer equipment
external contacts (fig 2).
It is necessary to read the following table for more information.
15
1
30
18
44
31
42
40
39
32
Terminal plug type DB 44
contacts male connector
Factory wiring
(Soldered-side view)
When the units containing the control circuits
are equipped with dry contacts outputs, it is the
customer’s responsability to use the outputs in
compliance with safety regulations.
CAUTION!
CAUTION!
Summary of Contents for Adixen ATH 1300 M
Page 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Page 24: ......
Page 26: ......
Page 32: ......
Page 44: ......
Page 50: ......
Page 62: ......
Page 64: ......
Page 68: ......
Page 70: ......
Page 82: ......
Page 88: ......
Page 96: ......
Page 98: ......
Page 112: ......
Page 114: ......
Page 116: ......
Page 122: ......