B 90
“Dry contacts“ connector wiring
1/1
GB 00739 - Edition 09 - Januar
y 01
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600 M/MT
Depending on the previous configuration the
contacts are closed (
see C 50
).
Signaling using outputs
Dry contact outputs
250 VAC - 3 A max
Controller outputs
12 VDC
Contact
functionnal status
Their functions are to copy
the data concerning the
pump operating status.
When the controller detects a bearing operating
fault, it opens the contact.
After a stop, it opens when the rotation speed
reaches 10000 rpm.
This contact can be used to control a secondary
isolation valve
which is used to maintain the
pressure in the chamber while the pump is reset
to atmospheric pressure.
0
1
0
1
ISV
Thermos.
Power off
Power on
Start on
Nominal speed
Alarm
De
Stop
Depending on temperature parameter
0
1
Water
valve
Depending on temperature parameter
(1)
Isol. valve
1 - 2
The contacts are opened or closed, depending on
the pump temperature and the selected temperature.
Thermostat
5 - 6
11-12
Air inlet*
9 - 10
The voltage delivered on this output depends on
the pump temperature and the selected
temperature.
Water valve
3 - 4
1
2
5
6
3
4
7
8
11 12
9
10
12 VDC
12 VDC
The voltage is delivered on this output during the
“Venting time” according to the configuration
(
see C 50
).
Air valve*
7 - 8
(1) This function is valid at 10000 rpm + setted delay when Stop or Ext. fault commands
are activated.
* Stop / Ext. fault functions allow to close the contacts at 10000 rpm + setted delay.
When the units containing the control circuits
are equipped with dry contacts outputs, it is the
customer’s responsability to use the outputs in
compliance with safety regulations.
CAUTION!
WARNING
HAZARDOUS VOLTAGE ENCLOSED
Danger risk of electric shock, disconnect main
power source and heater power. (if heater
is used, max 110/230 V, 300 VA, 50/60 Hz).
Prior to servicing and wait 5 minutes before
working on this equipment.
Summary of Contents for Adixen ATH 1300 M
Page 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Page 24: ......
Page 26: ......
Page 32: ......
Page 44: ......
Page 50: ......
Page 62: ......
Page 64: ......
Page 68: ......
Page 70: ......
Page 82: ......
Page 88: ......
Page 96: ......
Page 98: ......
Page 112: ......
Page 114: ......
Page 116: ......
Page 122: ......