Variation of the pump
rotational speed
The ATHM pump rotation speed can be selected and set
between a standby speed and the maximum speed.
This makes it possible to optimize pumping characteristics
according to each customer application (for example, high
pressure pumping).
A distinction is made between the following speeds:
-
reduced speed (STANDBY speed)
which can be set
between the low speed value and the maximum speed.
-
maximum speed
preselected at factory.
For an
inlet pressure
≤
10
-1
mbar,
the pump rotation speed
can be selected between :
- a minimum speed of
15000 rpm
and
- a maximum speed of
34000 rpm
for ATH 1300 M/MT
and
39000 rpm
for ATH 1600 M/MT
.