B 60
Water cooling connection
Characteristics
of water cooling
For ATH 1300 M
and ATH 1600 M
model
For ATH 1300 MT
and ATH 1600 MT
model
- Provide a water inlet
pipe and a tap to adjust
the flow rate.
- Connect the water valve
to the water inlet line using
a flexible tube following the
assembly diagram
(see A 60)
:
- Connect the other nipple
to the draining circuit.
Water
inlet
Water
outlet
Avoid to screw the connector on the valve inlet
port using a 13 mm flat wrench, to maintain it
(water leak risk).
CAUTION
In order to limit the corrosion and clogging of the cooling
pipes, it is recommended to use cooling water with the
following characteristics:
- treated soft water or non-corrosive industrial water
- pH between 7.5 and 11
- hardness < 7 milli-equivalent/dm
3
- Resistivity > 1500
Ω
.cm
- Solid pollution < 100 mg/dm
3
- Solid particle size (maxi): 0.03 mm
2
- Pressure range between 2 and 7 bars
- Temperature: 15 < T < 25°C
- Flow rate: 60 l/h
- Deionized water compatible
- Provide a water inlet
pipe and a tap to adjust
the flow rate.
- Connect the water inlet
line to one of the cooler
water fittings 1/4 NPT female on the pump, with the other
fitting connected to the water draining circuit via a tube
(supplied by customer).
Summary of Contents for Adixen ATH 1300 M
Page 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Page 24: ......
Page 26: ......
Page 32: ......
Page 44: ......
Page 50: ......
Page 62: ......
Page 64: ......
Page 68: ......
Page 70: ......
Page 82: ......
Page 88: ......
Page 96: ......
Page 98: ......
Page 112: ......
Page 114: ......
Page 116: ......
Page 122: ......