7
BenchCel Microplate Handling Workstation User Guide Addendum
7. Calculate the jog distance:
Jog distance = Z
initial
–
Z
adjusted
In the following example, the jog distance is –5.00 mm.
8. Calculate the adjusted Orientation sensor offset:
Adjusted Orientation sensor offset =
Initial Orientation sensor Jog distance
In the following example, the adjusted orientation sensor offset is
3.00 mm.
9. Type the adjusted orientation sensor offset in the
Labware
tab.
10. Select the next stack, and then click
Move to Sensor.
Check the Notch
Sensor readings. The notches should have much lower readings than
the corners. In addition, the difference between the notch readings
and corner readings should be at a maximum.
11. Repeat step 10 for the remaining stacks.
Z
initial
50.27 mm
Z
adjusted
55.27 mm
Jog distance
–5.00 mm
Initial Orientation sensor offset
8.00 mm
Jog distance
–5.00 mm
Adjusted Orientation sensor offset
3.00 mm
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...