Chapter 3: Setting up the BenchCel Workstation
BenchCel Microplate Handling Workstation R-Series User Guide
78
5.
In the graphical display area of the BenchCel Diagnostics dialog
box, click the stack that contains the spare microplates. In the
command menu that appears, click
Transfer to <teachpoint name>,
where <teachpoint name> is the name of the teachpoint you are
verifying. The robot picks up the first microplate in the stack and
moves it to the selected teachpoint.
6.
If the robot did not move the microplate to the correct location,
proceed to “Editing existing teachpoints” on page 79 to refine the
teachpoint.
7.
If the robot correctly placed the microplate at the teachpoint, move
the microplate back to the stack. To do this, in the graphical display
area of the BenchCel Diagnostics dialog box, click
Transfer to Stack
n
at the teachpoint. The robot moves the microplate from the
teachpoint back to the stack.
Click Transfer to <teachpoint name>.
Click Transfer to Stack n.
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...