Chapter 3: Setting up the BenchCel Workstation
BenchCel Microplate Handling Workstation R-Series User Guide
80
Locking the device
positions
After you finish setting the teachpoint and verified that it is accurate, you
can lock all the devices in their positions.
To lock the devices in their positions:
1.
Tighten the BenchCel integration plate clamp screw using the 3-mm
hex wrench.
2.
Check the teachpoints again.
Deleting teachpoints
To delete a teachpoint:
1.
In
BenchCel Diagnostics Profile
tab, verify that the correct teachpoint
file is loaded.
2.
In the graphical display area, double-click the teachpoint marker
that represents the teachpoint you want to delete. The
Teachpoint
Details
dialog box opens.
3.
Click
Delete.
The current teachpoint file will automatically be
updated when you delete a teachpoint.
4.
Save the device file.
Related topics
Tighten the screw to lock
the devices in position.
For information about
See...
Setup procedure
Integrating external devices
“Integrating the devices” on page 35
Moving between teachpoints
Summary of Contents for R Series
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Page 10: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Page 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Page 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Page 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Page 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Page 219: ......
Page 220: ...Agilent Technologies User Guide G5400 90001 ...