Making Swagelok Connections
A
Agilent Intuvo 9000 GC Installation
95
3
Push the tubing into the stainless steel plug (see
).
Figure 18
Assembling the fitting
4
Make sure that the front ferrule is touching the plug. Slide
the Swagelok nut over the ferrule and thread it onto the plug.
5
Push the tube fully into the plug, then withdraw it
approximately 1 to 2 mm (see
).
Figure 19
Insert the tubing
6
Finger-tighten the nut.
Nut and ferrules
Plug or
fitting held
in vise
Tubing
Front
ferrule
Back
ferrule
Nut
Insert
tubing fully
Withdraw tubing
1-2 mm
Tighten nut