Cabling Diagrams and Remote Start/Stop
B
Agilent Intuvo 9000 GC Installation
105
Number
Part number and description
1
G1530-60590, External events cable, 8-pin/spade lugs
G1580-87200, Caution label, Events cable, green in color
Connector
Signal name
Maximum rating
Wire color
Corresponds to valve #
24 volt control output
1
24 volt output 1
150 mA output
Yellow
5
2
24 volt output 2
150 mA output
Black
6
3
Ground
Red
4
Ground
White
Relay contact closures (normally open)
5
Contact closure 1
48V AC/DC, 250 mA
Orange
7
6
Contact closure 1
Green
7
7
Contact closure 2
48 V AC/DC, 250 mA
Brown or violet 8
8
Contact closure 2
Blue
8