Cabling Diagrams and Remote Start/Stop
B
Agilent Intuvo 9000 GC Installation
107
the 35900 A/D.
Figure 24
Analog output cable to an Agilent product
Remote start/stop cable, general use, 35900-60670
The pin assignments for the remote start/stop cable are listed in
.
4
Table 10
Remote start/stop cable connections
Connector 1, 9-pin male Connector 2, wire color Signal
1
Black
Digital ground
2
White
Prepare (low tone)
3
Red
Start (low tone)
4
Green
Start relay (closed during
start)
5
Brown
Start relay (closed during
start)
6
Blue
Open circuit
7
Orange
Ready (high true input)
35900-60670
1
6
5
9
Connector 1
Connector 2